| Display |
Meaning |
Action |
 |
There is a general failure of the thermometer’s function. |
Reset the thermometer by pushing the “ONE” and the “M” buttons simutaneously for 3 seconds. IF the thermometer still does not function, send the thermometer for repair. |
 |
The Thermometers own temperature is not within its operating range 16°C to 36°C. |
Put the thermometer in a room within the temperature range of 16°C to 36°C. Wait 30 minutes before taking a new measurement. |
 |
The measured temperature is not within the human temperature range of 34.6°C to 42°C. |
Repeat measurement and wait for beep to be heard before you remove the thermometer. |
 |
Battery voltage is low. Only a few more measurements can be taken. |
Replace the old battery with a new battery. |
 |
Battery voltage is low. No measurement is allowed. |
Immediately replace the old battery wtih a new battery. |
| No Touch does not function or respond. No display or fixed display. |
The thermometer should be reset or the battery should be replaced. |
Reset the thermometer by pushing the “ON” and the “M” buttons simultaneously for 3 seconds or replace the old battery with a new battery. If thermometer still does not function, end the thermometer for repair. |